MDS-150AAS24 BA

The Delta MDS-150AAS24 BA delivers 150 W at 24 V from a universal 90–264 Vac input, complete with low touch current (<0.1 mA normal, <0.3 mA single-fault) and 2×MOPP medical grade insulation. Certified to IEC 60601-1 & IEC 62368-1, it meets DOE Level VI & ErP Lot 7 efficiency requirements. Protection features include over-voltage, overload, temperature and short circuit with auto-recovery. Built for demanding medical, ITE, and industrial environments.

MDS-060BAS12 A

Features

  • Meet DOE Level VI and ErP Lot 7
  • Safety Approvals to IEC 60601-1 & IEC 62368-1
  • Compliant with IEC 60601-1-2 4th Ed. Requirements
  • Low touch current (<0.1 mA Normal & <0.3 mA single fault)
  • Over-Voltage/Load/Temperature & Short Circuit protections
  • 1 Million Hours MTBF
  • 2 × MOPP isolation

Key Technical Features

Input
Input Voltage
90 – 264 VAC
Input Frequency
47-63Hz
Output
# of Outputs
1
Output Voltage
12-24VDC
Current
6.25-10A
Power
150W
Efficiency
88%
Operational Temp
-10°C to +40°C
Dimensions
170 x 85 x 40 mm
Standards Compliance
IEC 60601-1 CB report, EN 60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 No. 60601-1, ANSI/AAMI ES 60601-1:2005/(R)2012